X-Ray Solution Xmark 6 Hardware technical parameters |
||||
H A R D W A R E |
X-RAY tube |
Tube type |
Sealed microfocus X-ray tube |
|
Tube voltage |
40-90kV |
|||
Tube current |
10-200uA |
|||
Focal spot size |
5-15μm |
|||
Cooling method |
Convection cooling |
|||
Detector |
Detector type |
HD digital flat panel detector (FPD) |
||
Imaging area |
130mm*130mm |
|||
Pixel matrix |
1536*1536 pixels |
|||
Inclined Angle |
0-60° |
|||
Inspection speed and accuracy |
Repeated test accuracy |
3μm |
||
Software inspection speed |
3.0s/point (Excluding loading and unloading time) |
|||
Table |
Standard size |
540mm*440mm |
||
Effective inspection area |
500mm×410mm |
|||
Load capacity |
≤5Kg |
|||
CNC programming |
Test parameters for different products can be stored in categories and called at any time. You can set the detection route or sequence of one or more products, and the program automatically completes the detection and stores the photos. | |||
Operating platform |
Mouse, keyboard, 2 operating modes | |||
Shell |
Inner lead plate | 5 mm thick lead plate (isolate radiation) | ||
Dimensions | 1360mm(L) * 1240mm(W) * 1700mm(H) | |||
Weight | 1250Kg | |||
Other parameters |
Computer | 24 inches Widescreen LCD/I3 CPU/2G Memory/200G hard disk | ||
Power supply | AC220V 10A | |||
Temperature and humidity | 22±3℃ 50%RH±10%RH | |||
Total power | 1700W | |||
Safety |
Radiation safety standard | Adopt steel-lead-steel protection structure. Any position 20mm from the shell, radiation≤1μSV/H, in line with international standards | ||
Safety interlock function | Two high-sensitivity limit switches are set in the door opening position for equipment maintenance. Once the door is opened, the X-ray tube will automatically power off immediately. | |||
Electromagnetic switch protection | The observation window has an electromagnetic switch, and the observation window cannot be opened when the X-ray tube is in working state. | |||
Observation window | With an observation window, the sample can be observed directly from the window while the machine is running. | |||
Emergency stop | The emergency stop is set in the prominent position of the operation console and the equipment body, can be pressed to quickly cut off the power supply system. | |||
X-ray tube automatic protection | Five minutes after the machine has no operation, the X-ray tube will automatically power off and enter the protection state. | |||
Automatic machine protection | Once any door or window of the machine is turned on, the machine immediately enters the shutdown protection state, and any operation cannot be performed. |
Hot
Xray inspection machine Xmark6
Xray inspection machine Xmark6
- X-ray tube and detector can move along Z direction, Speed of table moving along X-Y direction can be adjusted.
- High-definition digital flat panel detector.
- Max Inclined Angle is 65°, samples can be observed with a unique perspective
- Long life X-ray tube, maintenance free for life.
- Fault less than 2.5μm can be detected. High detection repeat accuracy.
Powerful - CNC Measuring Function, can test automatically, testing program can be edited.